Table 2 Pressures encountered at different flow rates across the three devices tested

From: Comparison of flow pressures in different 3-way infusion devices: an in-vitro study

 

Mean pressure (mmHg)

Flow rate (ml/hr)

3-way with anti-siphon valves

A

3-way without valves

B

Serial extension sets

C

10

178.7 ± 6

36.7 ± 1

32.7 ± 1

20

183.7 ± 7

37.0 ± 1

32.7 ± 1

30

185.7 ± 6

37.3 ± 1

32.7 ± 1

40

187.3 ± 5

37.3 ± 1

32.7 ± 1

50

189.7 ± 4

37.3 ± 1

32.7 ± 1

100

199.0 ± 1

38.7 ± 1

33.0 ± 1

400

243.3 ± 33

44.3 ± 1

35.7 ± 1