Skip to main content

Table 2 In-Vivo Laparoscopic Phase Results

From: Multiphase preclinical assessment of a novel device to locate unintentionally retained surgical sharps: a proof-of-concept study

  

X-ray + manual search (n = 9)

X-ray + MSF (n = 10)

MSF alone

(n = 33)

P value

Needle size

9 mm

4 (44.4%)

5 (50%)

16 (48.4%)

0.97

26 mm

5 (55.6%)

5 (50%)

17 (51.6%)

Needle present

 

7 (77.8%)

7 (70%)

21 (63.6%)

0.71

Needle absent

 

2 (22.2%)

3 (30%)

12 (36.3%)

Success: all

 

8/9 (88.9%)

10/10 (100%)

28/33 (84.5%)

0.42

Success if needle present

 

6/7 (85.7%)

7/7 (100%)

20/21 (95.2%)

0.49

Success if needle absent

 

2/2 (100%)

3/3 (100%)

8/12 (66.7%)

0.34

Time to determine (minutes)

 

2.2 ± 2.2

2.7 ± 2.1

2.8 ± 1.7

0.68

  1. There was no statistically significant outcome in comparing the three groups using this laparoscopic porcine model. The MSF was non-inferior to the X-ray groups in determining the presence of a needle.